page 21
This function allows the average load reading to be
displayed. The average starts being calculated when the
threshold (% of full-scale) is reached and stops being
calculated when the load falls back below this threshold.
To set AVERAGE over TIME, press and hold the MENU key
until page 1 of the main-menu appears. Using UP and
DOWN move the cursor key to AV/TIME. Press the ENTER
key.
The display will show AVERAGE/TIME OFF and SET.
Press ENTER to change AVERAGE/TIME OFF to AVERAGE
/TIME ON. Press DOWN to move the arrow cursor to SET
and press ENTER.
Define a % value of full-scale as the threshold value in the SET
T menu. Any load reading above this threshold will be
averaged over time with all the previous readings occurring
since the threshold value was exceeded. Averaging stops
when load readings fall back below the threshold.
Press ESC to return to main-menu page 1.
AVERAGE/TIME
Avg/Time sub-menu 1
Avg/Time sub-menu 2
25N
100N
75N
50N
0
Start Threshold
x Peak
TIME
Stop Threshold
Average