17. Emergency Stop Replacement
NOTE: General anti static precautions should be observed.
75. Depress the release lug on rear of
Emergency Stop button to release
loom connection - take care -
excessive force can break the lug.
76. Pull off loom connection.
77. Note the orientation of the switch,
and maintain this orientation on re-
assembly.
78. Using a suitable tool unscrew the
collar retaining the Emergency
Stop button and remove
completely, enabling the
Emergency Stop to be withdrawn
through the front of the front panel.
79. Fit replacement Emergency Stop.
Refitting the Emergency Stop button
is the reverse of the above procedure