EasyManua.ls Logo

MELAG MELAtherm 10 Evolution - 5 Process-Relevant Parameters (VRP); Universal-Program

MELAG MELAtherm 10 Evolution
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Process-relevant parameters (VRP)
23
5 Process-relevant parameters (VRP)
Universal-Program
Factory settings
Universal-Program
Pre-
cleaning
Cleaning Neutrali-
zation
Intermed.
rinsing
Disinfec-
tion
Drying
Partial cycle repetition 1 1 1 1 1 1
Holding time [sec.] 210 600 60 60 300 600
Temperature [°C] 22 55 -- -- 90 100
Metering temp. [°C] -- R: 40 -- -- K: 70 --
Metering conc. [ml/l] -- R: 6 N: 1.5 -- K: 0.3 --
DI proportion [%] 0 0 0 0 100 --
Quantity of water [l] 7 7 7 7 7 --
When changing the VRP, make a copy and enter the values in the following table:
Changes
Universal-Program
Pre-
cleaning
Cleaning Neutrali-
zation
Intermed.
rinsing
Disinfec-
tion
Drying
Partial cycle repetition
Holding time [sec.]
Temperature [°C] -- --
Metering temp. [°C] -- R: -- -- K: --
Metering conc. [ml/l] -- R: N: -- K: --
DI proportion [%] --
Quantity of water [l] --
Incorrect changes of the parameters can result in personal injury and/or damage to the device as well as
damage to property. MELAG does not accept any liability for such damage.
Date: ___________________ Signature service technician: ________________________________

Related product manuals