5 Process-relevant parameters (VRP)
24
Quick-Program
Factory settings
Quick-Program
Pre-
cleaning
Cleaning Neutrali-
zation
Intermed.
rinsing
Disinfec-
tion
Drying
Partial cycle repetition 1 1 1 1 1 1
Holding time [sec.] 90 420 60 30 300 600
Temperature [°C] 22 55 -- -- 90 100
Metering temp. [°C] -- R: 40 -- -- K: 70 --
Metering conc. [ml/l] -- R: 6 N: 1.5 -- K: 0.3 --
DI proportion [%] 0 0 0 0 100 --
Quantity of water [l] 7 7 7 7 7 --
When changing the VRP, make a copy and enter the values in the following table:
Changes
Quick-Program
Pre-
cleaning
Cleaning Neutrali-
zation
Intermed.
rinsing
Disinfec-
tion
Drying
Partial cycle repetition
Holding time [sec.]
Temperature [°C] -- --
Metering temp. [°C] -- R: -- -- K: --
Metering conc. [ml/l] -- R: N: -- K: --
DI proportion [%] --
Quantity of water [l] --
Incorrect changes of the parameters can result in personal injury and/or damage to the device as well as
damage to property. MELAG does not accept any liability for such damage.
Date: ___________________ Signature service technician: ________________________________