3 Setup and installation
12
3 Setup and installation
WARNING
Improper installation may lead to a short-circuit, fire, water damage or electrical shock.
This could result in serious injury.
n Only have the device set up, installed and started up by people authorized by MELAG.
u The device is not suitable for operation in explosive areas.
u The device is conceived for use outside patient surroundings. The device should be located a
minimum of 1.5m radius away from the treatment area.
u The documentation media (computer, MELAprint 42/44, MELAflash CF card reader) must be placed in
such a way that they cannot come into contact with liquids.
u Observe all the information contained in this chapter during commissioning.
Removing from the packaging
CAUTION
Danger of injury from incorrect carrying.
Lifting and carrying too heavy a load can result in spinal injury. Failure to comply with
these provisions can result in crushing.
n Comply with the safety regulations issued by your professional association.
The device and the floor unit are delivered on a europallet (WxDxH: 1200x800x144mm). Hold the
device by the reinforced areas of the device feet on the floor and the rear panel when carrying and lifting it.
Moving the device
The floor unit has three casters.
1. Slightly unscrew the fore caster for transport purposes.
2. After transportation, screw in the fore caster.