5 Process-relevant parameters (VRP)
26
Ophthalmo-Program
Factory settings
Ophthalmo-Program
Pre-
cleaning
Cleaning Neutrali-
zation
Intermed.
rinsing
Disinfec-
tion
Drying
I II
Partial cycle repetition 1 1 1 1 1 1 1
Holding time [sec.] 90 600 60 60 60 300 1800
Temperature [°C] 22 55 -- -- -- 90 100
Metering temp. [°C] -- R: 40 -- -- -- -- --
Metering conc. [ml/l] -- R: 6 N: 1.5 -- -- -- --
DI proportion [%] 0 0 0 100 100 100 --
Quantity of water [l] 7 7 7 7 7 7 --
When changing the VRP, make a copy and enter the values in the following table:
Changes
Ophthalmo-Program
Pre-
cleaning
Cleaning Neutrali-
zation
Intermed.
rinsing
Disinfec-
tion
Drying
I II
Partial cycle repetition
Holding time [sec.]
Temperature [°C] -- -- --
Metering temp. [°C] -- R: -- -- -- -- --
Metering conc. [ml/l] -- R: N: -- -- -- --
DI proportion [%] --
Quantity of water [l] --
Incorrect changes of the parameters can result in personal injury and/or damage to the device as well as
damage to property. MELAG does not accept any liability for such damage.
Date: ___________________ Signature service technician: ________________________________