2 Installation requirements
8
Property MELAtherm 10 Evolution DTA/MELAtherm 10 Evolution DTB
Installation location Dry and protected from dust, free of vibrations and shocks. There are no
installations in area of the door back.
Installation surface Level and horizontal Irregularities of the ground of max. 1cm can be offset by
setting the device feet.
Max. altitude 1500 m (depending on the set up height, it may be necessary to reduce the
disinfection temperature, see Exceptions when setting up at altitudes
[}page21])
Waste heat (with max.
solid load)
0.98 kWh (3.5 MJ)
Ambient temperature 5-40 °C (recommended max. 25 °C)
Relative humidity max. 80 % at temperatures up to 31 °C, max. 50 % at 40 °C (decreasing in a
linear fashion)
Max. floor loading
(normal operation)
Semi-integrated unit: Free standing: Top-frame device:
40.7 kg per device foot 42.7 kg per device foot 63.7 kg per device foot
Installation free-standing
NOTICE
In case of an incorrect installation of the device, there may be malfunctions during
operation.
The specified installation heights must be complied with.
Requirements to construction-side wastewater connection for a semi-integrated unit or free-
standing device
Connection height Remarks
0.4-1m ---
<0.4m Installation in a curve to at least 0.4m height
Fig.2: Connection scheme without floor unit
1 DI water connection 4 Connection for water outlet at the syphon
2 Cold water inlet 5 Type F (SCHUKO) socket
3 Dirt trap water inlet 6 CEE socket
PLEASE NOTE
Due to the length of mains lead and water pipe a sinistral connection is recommended.