SERVICE
3. Do not ship the power cord, serial cable, or any other accessories.
4. If the original packaging is not available, pack the box moderately tight with packing
material (e.g., Styrofoam
TM
peanuts or bubble wrap), ensuring the instrument is
suspended in the packing material.
5. On the RMA form, please verify the ship to and bill to information, contact name, and
problem description. If anything is incorrect, please contact Customer Support.
6. Tape the box in both directions for added support.
7. Include the RMA number in the attention line on the shipping label.
A Certificate of Calibration will be issued upon completion of the work.
4.5 TROUBLESHOOTING
Table 7 lists common problems and their solutions. If the problem is not listed or these
solutions do not solve the issue, contact Customer Support.
Table 7 Troubleshooting the AQUALAB
Problem Possible Solutions
AQUALAB does not turn on
Ensure power cord is securely attached to the back of the
instrument and plugged into the power outlet.
A power surge may have caused a fuse to blow. To change the fuses:
Unplug the power cord.
Locate the panel where the power cord plugs in. The fuse box is
on the right side of that panel. Press the release tab and pull the
fuse-holder out.
Pull the broken fuse out and replace with a 1.25-A, 250-V fuse.
CAUTION: Do not use any other kind of fuse to avoid risk of damaging the
instrument or voiding the warranty.
Replace the fuse holder and push it into the fuse well until the
release tab snaps in place.
Connect the power cord and turn the instrument on.
If the fuse blows again, a failed component may be causing the
problem. Contact Customer Support to make arrangements for
repairs.