EasyManua.ls Logo

METER GROUP AquaLab 4TE - Linear Offset

METER GROUP AquaLab 4TE
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46
SERVICE
Measure verification standard
Clean sample
chamber
Reread first
standard
Go to offset
procedure
Go to sampling
procedure
Read second
standard
Clean sample
chamber
Repeat process
Figure59 Verication standard owchart. A check indicates AQUALAB veried the standard within
specication; a cross indicates AQUALAB did not verify the standard within specications.
4.1.2 LINEAR OFFSET
After verification and cleaning, the instrument an offset may be necessary.
1. Navigate to the Configuration tab.
For the AQUALAB 4TEV, the currently selected sensor for the instrument will be the one
offset. Make sure to select the correct sensor before beginning the offset.
2. Select Calibration.
NOTE: The DUO model shows both water activity and moisture content on this screen. Select Water Activity.
3. Select Start (Figure60).
Figure60 Start dew point calibration
To return to the prior tab, select CANCEL.
4. A prompt will appear to insert a fresh standard and seal the chamber (Figure61).
NOTE: The same verification standard can be used to verify and adjust the linear offset. If using the same
verification standard, do not open the sample chamber between verification and offset.

Related product manuals