EasyManua.ls Logo

METER GROUP AquaLab 4TE - Restore Factory Defaults

METER GROUP AquaLab 4TE
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
51
AQUALAB
19. Measure a 0.76 a
w
verication standard according to the verication procedure detailed
in Section4.1.1.
If both verication readings are within ±0.015 a
w
, the instrument is ready to begin
testing.
If incorrect verication standard readings persist, contact Customer Support.
4.1.4 RESTORE FACTORY DEFAULTS
To restore original calibration settings, do the following:
1. Navigate to the Configuration tab using MENU.
NOTE: For TEV models, ensure the correct sensor is selected.
2. Select Calibration.
If using a DUO model, also select Water Activity.
3. Select Defaults to access the Restore Factory Defaults routine (Figure70).
Figure70 Calibration screen with Defaults selected
A conrmation dialog will appear (Figure71).
Figure71 Conrmation dialog to reset defaults
4. Select OK.
Select CANCEL to return to the Calibration menu.
The instrument will conrm the factory calibration has been restored (Figure72).

Related product manuals