18.10.2. Overloading the terminals
When external circuit connected to current output terminals is disconnected or there is higher voltage at the
load than permitted, the calibrator disconnects the output terminals and displays „Overload I output“ message.
The same message can be displayed when 50-turn coil is used for AC current output at frequencies above 80
Hz. It depends on the set current and the type of ammeter connected.
If the output terminals are disconnected due to time limitation of output current over 10 A, the calibrator displays
„Current timeout !“ message.
18.11. Generation of non-harmonic shapes
The multifunction calibrator can generate non-harmonic periodic signals with predefined shape. To allow the
setting of a non-harmonic output shape, the calibrator must be switched to AC U or AC I mode. In both cases,
an indication of the type of output shape (Shape xxxxx) is displayed under the frequency value. Press the
respective display button to change the shape of the output signal.
The calibrator can generate the following shapes:
SINE harmonic
PWM POS squarewave - positive, with adjustable duty cycle
PWM SYM squarewave - symmetrical, with adjustable duty cycle
PWM NEG squarewave - negative, with adjustable duty cycle
RAMP A ramp, symmetrical positive
RAMP B ramp, symmetrical negative
TRIANGLE triangular, symmetrical
LIM SINE harmonic with amplitude limitation (truncated sin)
Generation of non-harmonic signals has some limitations:
non-harmonic voltages can be generated in the 0.1 Hz to 1000 Hz frequency range
non-harmonic currents can be generated in the 0.1 Hz to 120 Hz frequency range
generation of these signals is limited to the voltage range up to 200 V and current range up to 2 A
non-harmonic signals cannot be generated in the P-E (power-energy) mode.
18.11.1. Control in the non-harmonic mode
Select AC voltage or AC current mode. The main section of the display shows the following data:
* main data of set current or voltage, unit of measurement
* relative deviation
* frequency
* selected SHAPE of the output signal
Keep pressing SHAPE display button to select the desired shape of the output signal:
The output terminals are automatically disconnected when changing the shape of the output signal or when
changing the relative deviation ∆%, if a non-zero relative deviation is set.

Table of Contents

Other manuals for Metrix CX 1651

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrix CX 1651 and is the answer not in the manual?

Metrix CX 1651 Specifications

General IconGeneral
BrandMetrix
ModelCX 1651
CategoryTest Equipment
LanguageEnglish

Related product manuals