135
Programming of values is done as follows:
Select the value to be changed by cursor buttons or the knob of the potentiometer.
Enter the numeric value using the units of measurement displayed in the header of the table.
Confirm the entry by pressing ENTER. New entry is copied to the respective position.
This way, all numeric entries in the table can be changed.
Press EXIT display button to return to the previous level when the programming is complete.
Note:
When the test program is executed, all steps are performed.
Within one test program, only one type of input signal and output signal can be used. It is impossible to switch
to other function of the calibrator in the middle of a program.
20.3.3. Setting the relays
SWITCH POLARITY and SWITCH ACTIVITY items of SETUP MENU are used to set-up the relay. The
following table lists the states of the relay depending on the values of SWITCH POLARITY, SWITCH ACTIVITY
and the test result.
SETUP MENU
RELAY BEFORE TEST
TEST RESULT
RELYA AFTER TEST
Switch polarity
OFF
disconnected
FAIL
disconnected
Switch activity
PASS
PASS
connected
Switch polarity
ON
connected
FAIL
connected
Switch activity
PASS
PASS
disconnected
Switch polarity
OFF
disconnected
FAIL
connected
Switch activity
FAIL
PASS
disconnected
Switch polarity
ON
connected
FAIL
disconnected
Switch activity
FAIL
PASS
connected