149
Capacitance test
Function
Range
Value (F)
Frequency (Hz)
Deviation
allowed(%value)
CAP
1 nF
1.00000e-09
1000
2.000
CAP
10 nF
1.000000e-8
1000
0.500
CAP
100 nF
1.000000e-7
1000
0.500
CAP
1 uF
1.000000e-6
500
0.500
CAP
10 uF
1.000000e-5
300
1.500
CAP
50 uF
5.000000e-5
300
2.000
Table XI
Frequency test
Function
Range
Value (Hz)
Frequency (Hz)
Deviation
allowed(%value)
FREQ
1 kHz
1000.0
0.005
Table XII
Multimeter test
Function
Range
Value (Hz)
Frequency (Hz)
Deviation
allowed(%value)
FREQ
1 kHz
1000.0 Hz
0.005
V-DC
10 V
10.0 V
0.013
V-DC
10 V
-10.0 V
0.013
A-DC
25 mA
0.019 A
0.017
A-DC
25 mA
-0.019 A
0.017
V-DC
20 mV
0.019 V
0.057
V-DC
20 mV
-0.019 V
0.057
V-DC
200 mV
0.19 V
0.024
V-DC
200 mV
-0.19 V
0.024
V-DC
2 V
1.9 V
0.020
V-DC
2 V
-1.9 V
0.020
O-4W
200 Ohm
100.0 Ohm
0.030
O-4W
2 kOhm
1000.0 Ohm
0.021
Table XIII