Technical Specifications
48 50,000-count Benchtop Multimeter
Technical Specifications
@ Note
Only assigned tolerance values or limits are guaranteed values. Non-
tolerance values are for information (NFC 42670 standard).
Please note: "n%L +nUR" means "n% of reading + n Representation Unit"
according to CEI 485).
Technical specifications are only guaranteed after a 30-minute warming up
period.
Measuring voltage
VAC and VAC+DC
“TRMS” alternating
voltage
@ Note
On this position, users can measure the rms value of AC voltage with its
direct component (no capacitive coupling) : "TRMS" measurement.
(
∗) Choose from menu
Results < 0.2 % are not displayed.
Admissible permanent maximum voltage: 600 V
ACrms
Specifications valid from 10 % to 100 % of F rate < 20 kHz
Specifications valid from 20 % to 100 % of F rate > 20 kHz
Influence of peak factor on V
precision:
• 0.5 % for peak factor between 2 and 3
• 1 % for peak factor between 3 and 6
Rejection: Common mode > 60 dB to 50 Hz or 60 Hz according to selection
Automatic or manual rate selection.
VDC direct voltage On this position, users can measure a direct voltage value or the direct
component of alternating voltage (you must keep to the range
corresponding to the rms value, see Serial mode).
(
∗) Choose from menu
Admissible permanent maximum voltage: 600 V
DCrms
Rejections: Common mode > 120 dB to 50 Hz and 60 Hz according to
selection
Serial mode > 60 dB to 50 Hz and 60 Hz with active filter
Automatic or manual range selection.
Range Precision Resolution Input resistance Admissible
overload
500 mV 0.08% + 3D 0.01 mV
10 MΩ / (∗)1 GΩ
5 V 0.08% + 3D 0.1 mV
11 MΩ
50 V 0.08% + 3D 1.0 mV
10 MΩ
500 V 0.1% + 3D 10 mV
10 MΩ
600 V 0.1% + 3D 0.1 V
10 MΩ
Accuracy
Rate
Input
resis-
tance
1 kHz
4 kHz
4 kHz to
50 kHz
AC
10 MΩ /
(∗) 1 GΩ
10 µV 0.5 %
+ 40 D
- - - - -
5 VAC
11 MΩ
0.1 mV
50 VAC
10 MΩ
1.0 mV
AC
10 MΩ
10 mV
3 %
+ 30 D
5 %
7 %
+ 30 D
AC
10 MΩ
0.1 V
(1 min
max)
1000 V
DC
or
700 V
AC
0.5 %
+ 30 D
1 %
2 %
+ 30 D
- - -
DC
or
700 V
AC
(1 min max.)