EasyManua.ls Logo

Metrix MTX 3250 - Resistance Measurement Specifications; Continuity Mode Specifications; Diode Test Specifications

Metrix MTX 3250
133 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Specifications
50 50,000-count Benchtop Multimeter
Technical Specifications
Measuring
resistance
( )
@ Note
On this position, users can measure a resistance value.
(∗) with default-activated filter
Automatic or manual range selection.
Open circuit maximum voltage: 9 V.
A shielded cable is recommended for measurements exceeding 5 M.
Continuity mode
( )
On this position, users can measure a resistance value until 500 , with
continuous audible indication at 4 kHz.
Continiuty mode detection threshold 10 to 15 (reply time 1ms).
Open circuit maximum voltage: 9 V.
Diode test ( )
Indicating interface voltage in current direction from 0 to 4.5 V for only
one range
Precision: 0.2 % ± 3 D
Indicating overshoot in reverse direction
Measurement current: about 1 mA - V 2 V
Open circuit maximum voltage: 9 V
Rate Precision
Measurement
currrent
Resolution Protection
0 to 500
0.1% + 5D 1 mA
100 m
5 k
0.1% + 3D 100 µA
1
50 k
0.1% + 3D 10 µA
10
500 k
0.1% + 3D 1 µA
100
5 M (∗)
0.5% + 3D 100 nA
1 k
50 M (∗)
1% + 5D 50 nA
10 k
600 Vrms
Rate Precision
Measurement
current
Resolution Protection
Beeper 0.1% + 5D about 1 mA 100 m 600 Vrms
(cont.)