4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
88
.GasFlow: Average gas flow during sample heating time.
.UnitFlow: Unit of gas flow.
4.2.2.74. Info.ActualInfo.Display.L1 up to 32 ASCII characters
Info.ActualInfo.Display.L8
up to 32 ASCII characters
Info.ActualInfo.Display.DelAll $G
Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
1. Lock the display, see 4.2.2.90.
2.
Delete the whole di
splay (.DelAll).
3.
For writing onto the display, the
standard character set will be used.
4. Unlock the display, see 4.2.2.90
5.
Delete the whole di
splay (.DelAll).
6.
Send a value to nod &Config.Aux.Re
sDisplay (see 4.2.2.47) to refresh the
display.
$Q sends the contents of the corresponding display line.
4.2.2.75. Info.ActualInfo.Comport.N
umber read only
$Q sends the comport number of the Coulometer where the PC i
s connected.
4.2.2.76. Info.Assembly.C
ycleTime read only
I
nfo.Assembly.ExV read only
I
nfo.Assembly.Devi
c
eTemp read only
Inquirie
s regarding basic variables of the assembly.
.Cycle time: Time of measuring cycles in s (0.4).
.ExV: Volume of the Dosing Unit of the connected Dosino in mL.
.DeviceTemp: Internal temperature of Coulometer in °C.
4.2.2.77. Assembly.GenEl.Pulse $G
Assembly.GenEl.Pulse.Length 0...2000
A
ssembly.GenEl.Pulse.Current
0, 100, 200, 400
Control of the generator electrode. The pulse will be generated with
&A.G.P$G.
.Length: Length of pulse in 2000 steps. 2000 means a pulse of 400 ms
(e.g. a pulse of 150 ms would mean 750 steps).
.Current: Current for pulse in mA.
4.2.2.78. Assembly.Meas.Status ON, OFF
Assembly.Meas.Ipol 2, 10, 20, 40
Control of the i
ndicator electrode. When the measuring function is switched
on, no method can be started at the Coulometer.
.Ipol: Polarization current in uA.
4.2.2.79. Assembly.Outputs.AutoEOD
ON, OFF
Assembly.Outputs.SetLines $G
A
ssembly.Outputs.SetLi
nes.L0 active, inactive, pulse, OFF
up to .L13