7.1 Statistical function
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
916 Ti-Touch
7 Performing titration with supplementary func-
tions
7.1 Statistical function
The statistical function can calculate the mean value and the absolute and
relative standard deviations (s abs and s rel) from several determinations.
The calculation is carried out continuously, i.e. the calculation is updated
after every determination which is carried out until the defined number of
determinations has been reached.
Several results can be calculated for one determination, all of which can
be incorporated in the statistical calculation.
Activating the statistics
1
Edit the method options
In the main dialog, tap on [Edit parameters].
Tap on [Method options].
Here you can define the statistical function for this method. You
can specify the number of determinations for which the respective
statistical calculations are to be carried out.
These settings apply for all determinations which are carried out
with this method.
Tap on the check box Statistics.
Enter a value under Number of determinations.
Tap on [
].

Other manuals for Metrohm 916 Ti-Touch

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 916 Ti-Touch and is the answer not in the manual?

Metrohm 916 Ti-Touch Specifications

General IconGeneral
BrandMetrohm
Model916 Ti-Touch
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals