EasyManuals Logo

Metrohm 916 Ti-Touch User Manual

Metrohm 916 Ti-Touch
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
■■■■■■■■■■■■■■■■■■■■■■
7 Performing titration with supplementary functions
916 Ti-Touch
■■■■■■■■
45
Select the CALC command and tap on [Edit command].
The second calculation (R2) should not be incorporated in the
statistical calculation.
Select R2 and tap on [Edit].
Tap on [Result options].
The variable name under which the calculated mean value is to be
stored is specified under Variable for mean value. You can
change this variable assignment in order to switch off the statisti-
cal calculation for this result.
Tap on the selection symbol next to SMN2 and select off.
Return to the main dialog with [
].
7.2 Saving the determination and the PC/LIMS report
You can have the determination data (measuring point list, results, etc.) of
a titration which has been carried out saved automatically. This makes it
possible to reprocess or print out this data later. You can save the deter-
minations to a USB storage medium or to a drive in a computer network.
If you would like to administer the determination data in a database on a
PC, then you can save it in a report called PC/LIMS report or transfer it
directly to a PC through a serial port. The tiBase database from Metrohm
is available to you for administering the determination data on the PC.
Defining the automatic saving
Proceed as follows:
1
Open the method options
In the main dialog, tap on [Edit parameters].
Tap on [Method options].

Other manuals for Metrohm 916 Ti-Touch

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 916 Ti-Touch and is the answer not in the manual?

Metrohm 916 Ti-Touch Specifications

General IconGeneral
BrandMetrohm
Model916 Ti-Touch
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals