4 Installation
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
■■■■■■■■
27
■ Adsorption cartridge holder (6.2057.080) with two clips.
■ CO
2
adsorption cartridge CW (6.2837.100)
All of the parts are located in one accessory kit: Vario/Flex SeS
(6.5000.020).
Preparing the CO
2
adsorption cartridge CW
Prepare the CO
2
adsorption cartridge CW for use as follows:
1
Remove the protective cap from the tip of the CO
2
adsorption car-
tridge CW.
2
Attach the dust filter to the tip of the CO
2
adsorption cartridge CW.
3
Remove the label from the lid of the CO
2
adsorption cartridge CW.
This uncovers the small opening in the lid of the CO
2
adsorption car-
tridge CW, through which air is aspirated.
The CO
2
adsorption cartridge CW is now ready for installation.
NOTE
The new CO
2
adsorption cartridge CW (6.2837.100) works without
attached H
2
O adsorption cartridge.