4.20 Initial start-up
■■■■■■■■■■■■■■■■■■■■■■
32
■■■■■■■■
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
Connecting the power cord
Accessories Power cord, three-core with IEC 60320 instrument plug type C13. Con-
ductor cross-section 1 mm
2
/ 18 AWG. Power plug according to customer
requirement (6.2122.XX0).
Do not use a not permitted power cord.
1
Plugging in the power cord
■ Plug the power cord into the instrument's power socket.
■ Connect the power cord to the power grid.
4.20 Initial start-up
Even before the guard column and separation column are installed, the
entire system must be completely rinsed with eluent for the first time.
Rinsing the IC system
CAUTION
The separation column and the guard column are not permitted to be
installed at the time of the initial start-up.
Make sure that a coupling (6.2744.040) is being used instead of the
columns.
1
Preparing the software
■ Start the MagIC Net computer program.
■ Open the Equilibration tab in MagIC Net: Work-
place ▶ Run ▶ Equilibration.
■ Import (or create) a suitable method.
Also see: MagIC Net Tutorial and online help.
2
Preparing the instrument
■ Ensure that the eluent aspiration tubing is immersed in the eluent
and that there is enough eluent in the eluent bottle.