26
■■■■■■■■
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
4.14 Metrohm CO
2
Suppressor (MCS)
4.14.1 Connecting the MCS
The MCS is connected between the Metrohm Suppressor Module (MSM)
and the conductivity detector.
Connecting the MCS
Figure 6 Connecting the MCS
1
Air aspiration capillary
For drawing in air with low CO
2
content (via
the CO
2
adsorption cartridge).
2
Pressure screw, short (6.2744.070)
Installed on the air aspiration capillary.
3
Luer coupling (6.2744.120)
Mounted on the air aspiration capillary with
a pressure screw (6.2744.070).
CAUTION
If the MCS is not used, the inlet and outlet must be sealed with the
threaded stoppers (6.2744.220).
4.14.2 Installing adsorption cartridges
The aspirated air has to have the lowest CO
2
content possible in order for
the CO
2
to be removed from the eluent efficiently. In order to achieve this,
the air is aspirated through a CO
2
adsorption cartridge CW (6.2837.100).
Accessories
For this step, you need the following accessories: