4 Installation
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
■■■■■■■■
35
Connecting the guard column
Rinsing the guard column
1
Rinsing the guard column
Place a beaker under the guard column's outlet.
Start manual control in MagIC Net and select the high-pressure
pump: Manual Manual control Pump
Flow: in accordance with column leaflet
On
Rinse the guard column with eluent for approx. 5 minutes.
Stop the high-pressure pump in the manual control in MagIC Net
again: Off.
4.22 Connecting the separation column
CAUTION
New separation columns are filled with a solution and sealed with stop-
pers on both sides. Before inserting the column, ensure that this solu-
tion can be mixed with the eluent being used (follow the information
provided by the manufacturer).
NOTE
Connect the separation column only after the initial start-up of the
instrument. Until that point, insert a coupling (6.2744.040) instead of
the guard column and separation column.

Table of Contents

Other manuals for Metrohm 930 Compact IC Flex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 930 Compact IC Flex and is the answer not in the manual?

Metrohm 930 Compact IC Flex Specifications

General IconGeneral
BrandMetrohm
Model930 Compact IC Flex
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals