Page #44 background imageLoading...
Page #44 background image
4.22 Connecting the separation column
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
Connecting the separation column
1
Rinsing the separation column
Place a beaker under the outlet of the separation column.
Start manual control in MagIC Net and select the high-pressure
pump: Manual Manual control Pump
Flow: Increase gradually up to the flow rate recommended
in the column leaflet.
On
Rinse the separation column with eluent for approx. 10 minutes.

Table of Contents