4 Installation
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
■■■■■■■■
33
■ – STREAM method (recommended): Ensure that the aspira-
tion tubing for the regeneration solution is immersed in the
solution and that there is enough solution. Check whether
the detector outlet capillary is connected to the Metrohm
Suppressor Module (MSM)'s inlet capillary for rinsing solu-
tion (labeled rinsing solution).
– Alternatively, if two bottles are being used for rinsing and
regeneration: Ensure that the aspiration tubings for the aux-
iliary solutions (regeneration solution and rinsing solution)
are immersed in the respective solutions and that there is
enough solution in both bottles. Check whether the detec-
tor outlet capillary is guided into the waste container or is
connected with the waste collector.
■ Switch on the instrument.
MagIC Net detects the instrument and all of its modules.
3
Starting equilibration
■ Start the equilibration in MagIC Net: Workplace ▶ Run ▶ Equili-
bration ▶ Start HW.
4
Deaerating the high-pressure pump