Select Configure, Options from the menu bar.
Verify that these selections are clicked:
Instrument must stabilize before data
acquisition
Use Auto-Linearization
Reference Standardization
Use Instrument Calibration
These selections are explained in more detail,
further in this manual. Because some selections do
not apply to the MultiVial, they are not selected.
Reference Standardization and Instrument
Calibration must be selected for transferability
between similar instruments.
Click “OK” to exit the dialog box.
Select Configure, Tray from the menu bar.