1 INSTALLATION OVERVIEW
This document contains procedures to install and verify the operation of the Gemini VII 2390
analyzer (models 2390a, 2390p, and 2390t).
Use these installation instructions and checklists to ensure that installation procedures are
completed properly.
n If a Micromeritics Service Technician performs this installation, additional charges
apply. See Contact Us on pageiv for information on how to contact Micro-
meritics.
n It is recommended that installation procedures be performed in the order presented
in this document.
n Use this installation instruction manual in conjunction with the analyzer operator
manual.
n This manual contains instructions for installations in standard and 21CFR11
environments. For instructions on installing the software in a 21CFR11
environment, see the Confirm Administrator Guide [part number 004-42821-01].
This document can be found on the Micromeritics web page
(www.Micromeritics.com).
PREPARE FOR INSTALLATION
Before installation, ensure the operator has reviewed thoroughly the analyzer Pre-installation
Instructions and Checklist. Ensure all required equipment and supplies are unpacked and the
required personnel are available.
When the equipment is received, unpack and inspect the contents of the shipping container(s).
Use the packing list to verify that all products, accessories, software (if applicable), and
documentation are received intact and in the correct quantity. The shipping container(s) and
contents should be inspected within a few days of receipt in the event damage or loss has
occurred. Sort through all packing material before declaring missing equipment or parts.
Micromeritics recommends saving all shipping containers until installation of the equip-
ment is complete. All shipping containers where equipment is to be declared as dam-
aged or lost must be examined by the claims investigator prior to completion of the
inspection report.