Helium Gas Line Rev [n] if the helium cylinder is being verified.
2. Select Report after test and select Preview as the destination.Resize the window if this
option does not display.
3. Click Start. From the View drop-down list, select either Operation, Instrument Log, or Instru-
ment Schematic.
4. A series of prompts display requiring operator response. The length of time a test will run is
also indicated.
5. A popup window indicates the test is complete. Click OK.
6. Click the Instrument to Regulator Shut-off Valve -1 tab to display its report.
The Leak Rate field should read Passed indicating the gas line from the analyzer to the
regulator is clean and leak-free.
If the Leak Rate field indicates Failed, a leak is indicated. Check the connections from the
analyzer to the gas regulator valve. Tighten as necessary, then run the test again.
7. Click the Instrument to Gas Bottle Shut-off Valve - 2 tab to display its report.
The Leak Rate field should read Passed indicating the gas line from the analyzer to the
regulator is clean and leak-free.