<Command> :MEASure:BIN:PARAmeter
{OFF|RDC|LS|LP|CS|CP|Q|D|RS|RP|Z|DEG|RAD|R|X|Y|G|B|E|U}
:MEASure:BIN:PARAmeter?
FunctionIt sets or queries the BIN Parameter.
Description
Set syntax
:MEASure:BIN:PARAmeter Z(only allow the parameter which is being used
under
the meter mode.)
Query syntax
:MEASure:PARAmeter?
Return
dataZ(Format is in <disc>)
<Command> :MEASure:BIN:NUMBer {2|3|4|5|6|7|8|9|MAXium|MINimum}
:MEASure:BIN:NUMBer?
FunctionIt sets or queries the BIN NUMBER.
Description
Set parameter
The value of bin number is {2~9|MAXimum|MINimum}.
Set syntax
:MEASure:NUMBer 4
Query syntax
:MEASure:NUMBer?
Return
data4(Format is in <NR1>)
<Command> :MEASure:BIN:METHod {EQUal|SEQuential|TOLerance|RANDom|0|1|2|3}
:MEASure:BIN:METHod?
FunctionIt sets or queries the BIN METHOD.
Description
Set parameter
EQUal,0| SEQuential,1|TOLerance,2 |RANDom,3
Set syntax
:MEASure:METHod SEQ
Query syntax
:MEASure:METHod?
Return dataSEQ(Format is in <disc>)
<Command> :MEASure:BIN:MODE <ABSolute|DEViation|PERCent|0|1|2>
:MEASure:BIN:MODE?
FunctionIt sets or queries the BIN MODE.
Description
Set parameter
ABSolute,0|DEViation,1|PERCent,2
Set syntax
:MEASure:BIN:MODE DEViation
Query syntax
:MEASure:BIN:MODE:MODE?
Return dataDEV(Format is in <disc>)
When the setup is 0, it returns ABS , When the setup is 1, it returns DEV , When the setup is
2, it returns PERC.
<Command> :MEASure:BIN:NOMinal <nominal value NR3 >
:MEASure:BIN:NOMinal?
FunctionIt sets or queries the BIN NOMINAL.
Description
Set syntax
:MEASure:BIN:NOMinal 1000
:MEASure:BIN:NOMinal 1K
:MEASure:BIN:NOMinal 1E+03
Query syntax
:MEASure:BIN:NOMinal?
Return data 1.000000E+03 (Format is in <NR3>)