<Command> :MEASure:BIN:LIMit <nominal value NR3 >
:MEASure:BIN:LIMit?
Function:It sets or queries the BIN LIMIT.
Description:
Set syntax
:MEASure:BIN:LIMit 0.001,100M,1k,1000k
Query syntax
:MEASure:BIN:LIMit?
Return
data +1.000000E-03, +1.000000E-01, +1.000000E+03, +1.000000E+06 (Format is in
<NR3>)
<Command> :MEASure:FILE:LOAD <filename>
Function:It opens the meter mode's file.
Description:
Set syntax
:MEASure:FILE:LOAD MICROTEST.( load the file “MICROTEST”)
<Command> :MEASure:FILE:LOAD?
Function:It queries the meter mode's filename which is being using.
Description:
Query syntax:
MEASure:FILE:LOAD?
Return
dataMICROTEST
<Command> : MEASure:BIAS
:VOLTage<NR2>
: MEASure:BIAS:VOLTage?
Set parameter: The value of BIAS Voltage -12.000~12.000
Function:It sets or queries the BAIS Voltage.
Description:
Set syntax:MEASure:BIAS
:VOLTage -12
Query syntax: MEASure:BIAS
:VOLTage?
Return
data -1.200000E+01
<Command> : MEASure:BIAS:STATe<ON|OFF|0|1>
: MEASure:BIAS:STATe?
Function:It sets or queries the BIAS Voltage enabled/disabled.
Description:
Set syntax:: MEASure:BIAS:STATe
1
Query syntax: MEASure:BIAS:STATe
?
Return
data When the setup is OFF, it returns 0.
When the setup is ON, it returns 1.
Function:
It queries all of the meter mode's filenames that are stored in the memory.
Description:
Query syntax: MEASure:FILE?
Return
data
N, filename1, filename2, filename3, ..., filenameN
where N= total file count
<Command> : MEASure:STATistic {ON|OFF|0|1}
: MEASure:STATistic?
Function:
It sets or queries the statistic function in the meter mode.
Description:
Set parameter {ON|OFF|0|1}
Set syntax: MEASure:STATistic 1
Query syntax: MEASure:STATistic?
Return
data
0|1
where 0 = statistic function is disabled.
1= statistic function is enabled.