65
<Command> :CORRection:OPEN
:CORRection:OPEN?
Function:It sets or queries to do open correction.
Description:
Set syntax
:
CORRection:OPEN
Query syntax
:CORRection:OPEN?
Return
data When correction fail, it returns 0.
When correction passes, it returns 1.
<Command> :CORRection:SHORt
:CORRection:SHORt?
Function:It sets or queries to do short correction.
Description:
Set syntax
:
CORRection:SHORt
Query syntax
:CORRection:SHORt?
Return
data When correction fail, it returns 0.
When correction passes, it returns 1.
<Command> :CORRection:OPEN:STATe {OFF|ON|0|1}
:CORRection:OPEN:STATe?
Function:It sets or queries the status of open correction.
Description:
Set syntax:CORRection:OPEN:STATe ON
Query syntax
:CORRection:OPEN:STATe?
Return
dataWhen the setup is OFF, it returns 0.
When the setup is ON, it returns 1.
<Command> :CORRection:SHORt:STATe {OFF|ON|0|1}
:CORRection:SHORt:STATe?
Function:It sets or queries the status of short correction.
Description:
Set syntax:CORRection:SHORt:STATe ON
Query syntax
:CORRection:SHORt:STATe?
Return
dataWhen the setup is OFF, it returns 0.
When the setup is ON, it returns 1.
<Command> :CORRection:CABLe {0|0.5|1|2}
:CORRection:CABLe?
Function:It sets or queries the status of cable length correction.
Description:
Set syntax:CORRection:CABLe 1
Query syntax
:CORRection:CABLe?
Return
data0|0.5|1|2