3.1.2 RAM Configuration Table
This item allows the viewing of the 8751/8761 machine and version information; it lists the machine’s
model number, test pins, total space used by files and the total remaining space etc. And it can perform
system update.
(1) Press [Func]; please use [↑] or [↓] to select RAM configuration table and then press [Enter].
(2) The LCD screen will display as follows:
3.1.3 Test Parameter Calibration
This function includes DC high voltage calibration, AC high voltage calibration and internal resistance
offset.
(1) Press [Func].
(2) Please use [↑] or [↓] to select test parameter calibration and then press [Enter].
(3) The LCD screen will display as follows: