English - 26
003-2915-99
© Midmark Corporation 2018
Cycle Operation - continued...
Note
On units using the metric display,°F will display as°C and PSI will display as KPA.
Pressing enables operator to change DRY time from 0 to 60 minutes in 1 minute
increments on a pre-set cycle.
Pressing decreases time. Pressing increases time.
If the new Drying Time is desired on a regular basis pressing “P” after adjusting the dry time will
store the new time as the default dry time for the selected cycle.
WARNING
STOP button may be pressed at any time to stop or interrupt
a cycle. Goods must not be considered sterile if this occurs
before the Dry Cycle. Sterilizer will return to SELECT CYCLE mode.
Step 3: Press ‘Start’ button to initiate cycle.
You will hear a “beep” for two seconds, indicating the cycle has started.
Start
Stop