Cycle Operation - continued...

The chart below illustrates the display messages that will appear during each stage of the cycle.
Stage of Cycle Description Display
Filling 
When water reaches the proper level...
Heating Display changes as temperature and
pressure in chamber changes.
(Metric units may be displayed if desired).
Sterilizing Sterilizing begins when correct
temperature / pressure is reached.
Time remaining in cycle counts down
while current temperature / pressure in
chamber is continuously updated.
Ready to Vent ‘READY TO VENT’ is displayed when
10 seconds remain in sterilization cycle.
Fast Vent 
the vent valve opens. Steam / water are
released back into the reservoir.
The display changes as temperature and
pressure in chamber changes.
Caution
Keep clear when M9 / M11 door is ready to open!
Failure to do so could result in severe burns from steam being released.
Door To Open
(automatic)
Pertains only to
M9/M11 Steam Sterilizers

that the door is about to open.

the door actuates to partially open (drying
mode) position.
This chart continues on the following page...