Washing without dispenser compartment
26
Dispenser calibration
Dispenser pump 1
0/1
0 st:
P: No
V: 100 ml
Press multifunction button0/1.
Stop the pump by pressing button0/1, or the pump will stop auto-
matically after 60seconds.
Measure the required amount in the measuring vessel.
Dispenser calibration
Dispenser pump 1
0/1
30 st: P: 200 ml/minV: 100 ml
Enter the amount measured in field"V:".
The throughputP is calculated in ml/min. by the controls.
Press the "Save" button.
The measured value is assigned to the pump.
Reattach the hose.