99
Dispenser calibration
Dispenser pump 1
- OK +
V:
100 ml
0/1
30 st: P: 200,0 ml/min
Enter the amount measured in field"V:".
The throughputP is calculated in ml/min. by the controls.
Press the "Save" button.
The measured value is assigned to the pump.
Reattach the hose.
Dispensing throughput (D1 to D13)
The throughput of the connected dispensing pumps can be set from
0ml/min to 9999ml/min in 1ml/min steps.
Access the following content:
Supervisor level
Settings
Dispensing throughput
Dispensing correction factor
OK
Select "Dispensing throughput".
The following will appear on the display:
Settings
Dispensing throughput
Dispenser pump 1
Dispenser pump 2
OK
: 100,0 ml/min
100,0 ml/min:
Options:
No
1to9999.0ml/min