This area is used for the display of different messages:
• Possible actions using
and buttons
• Information and instructions regarding test method step (in Test Method Mode)
• Alert messages and error messages
Filtration Parameters
Timer
The timer function stops the pump automatically after a preset filtration time.
The timer value can vary from off (not used) to 999 (see Using the Pump in
Standard Mode).
NOTE
The timer default value is off.
Speed
The pump speed value can vary from 1 to 150.
NOTE
The default speed value is 75.
Pressure
The Steritest
®
Symbio Pump constantly monitors the pressure inside each
Steritest
®
canister and indicates when pressure goes above normal working
conditions. Colored bars and audible beeps indicate the pressure level. A white
bar indicates that the pump head is not running.