The liquid is transferred to the tested canister, and the pressure on the
pressure gauge increases.
CAUTION
If the pressure reaches 53 psi (3.6 bar) and the pump does not
automatically stop, press the control knob to abort the test. See Pressure
Sensors Test Issues, Overpressure.
NOTE
If no pressure is detected after about 30 seconds, the test stops. See
Pressure Sensors Test Issues, No Pressure Increase Detected.
11. As soon as the pump stops, read the pressure on the pressure gauge. The
pressure should be between 37 psi (2.5 bar) and 53 psi (3.6 bar).
NOTE
If the pressure is not between 37 psi (2.5 bar) and 53 psi (3.6 bar). See
Pressure Sensors Test Issues, Pressure Outside The Acceptable
Range.