7
Pre Installation
Installation Requirements
Feedwater
Requirements
Type of Feedwater Elix, RO, distilled or DI water
Flowrate
1.5 LPM (Litres per minute)
Minimum Pressure > 0 bar (Feed from a tank)
Maximum Pressure
0.5 bar
Temperature
5 to 35 °C
Conductivity
< 100 µS/cm
TOC < 50 ppb
NOTE:
If the feedwater pressure is greater than 0.5 bar, then the Pressure Regulator
(ZFMQ000PR) is required.
Feedwater
Connection Needed
Feedwater Piping Connection 1/2 inch Male GAZ, NPTM or BSPM
Environmental
Requirements
Indoor Use Only
Storage Temperature
5 °C < T < 40 °C
Operating Temperature
5 °C < T < 40 °C
Relative Humidity
Should not exceed 80% for temperatures below 31 °C
Should stay within 50% to 80% between 31 °C and 40 °C.
Altitude < 3000 metres
Installation Category II
Pollution Degree 2
Optional Equipment You May Need
Support and Bracket
The Millipore SAS Catalogue Number for:
The Wall Mounting Bracket is WMBSMT001.
The Support for remote dispenser is RDHSTAND1
The Spacer for remote dispenser is RDHSPACE1
The Small mast for remote dispenser is RDHSMAST1
The mounting hardware for attaching the bracket to a wall is not included and must be
supplied.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Millipore Synergy R and is the answer not in the manual?

Millipore Synergy R Specifications

General IconGeneral
BrandMillipore
ModelSynergy R
CategoryWater Filtration Systems
LanguageEnglish

Related product manuals