Page #30 background imageLoading...
Page #30 background image
Using the Synergy R System
24
How to Dispense an Exact Amount of Product Water (F01)
1
Press the Main and “-Buttons together
to enter the menu. The Display will show
F01.
2
Press the “+” or “- Button to adjust the
exact amount of Product Water (in Litres)
needed. Pre-set volumes of water can be
adjusted from 0.25 L to 9.75 L in
0.25 increments.
NOTE:
The accuracy of this is dependent
upon the accuracy of the Flow
Calibration performed in Section
How to Calibrate the Flowrate from the
Synerg
y R System (F02).
3
Press the Dispensing Button once. The
system will dispense water. The Display
will show the amount of water dispensed
and the Product resistivity.
NOTE:
To stop dispensing water, press the Dispensing Button once again.
4
To exit the menu, press and hold the Main
Button for 2 seconds.
To display the next menu option, press the
Main Button once.

Table of Contents