29
Maintenance
Maintenance Schedule
In order to maintain good performance of your water system respect the maintenance schedule and replace
consumables as required.
What to do When? How to?
SynergyPak
cartridge
Replacement
When the Pack
Alarm display is
blinking.
See Section How
to Replace the
SynergyPak
cartridge
When the
system resistivity
display is
blinking.
Final Filter
Replacement
The Final Filter is replaced when the
SynergyPak cartridge is replaced or when
the Product Water flowrate drops.
See Section How
to Replace the
Final Filter
Flow
Calibration
When the Final Filter is replaced, or
periodically.
See Section How
to Calibrate the
Flowrate from the
Synergy R System
UV Lamp
Replacement
and UV Timer
Reset
When the UV
Lamp Alarm
display is
blinking.
See Section How
to Replace the UV
Lamp (UV System
only) and How to
View or Reset the
UV Lamp Timer
(C03)
Screen Filter
Cleaning
2 times a year or as necessary.
See Section How to
Clean the Screen
Filter