Maintenance
35
How to Replace the UV Lamp (UV System only)
The red UV Lamp Alarm will be blinking on the Display when it is time to exchange the UV Lamp. The message is
shown when the UV Timer has reached 0 days
(
see
Section How to View or Reset the UV Lamp Timer (C03)).
HAZARD
No electrical power should be going to the system at this time. Accidental
exposure to ultraviolet light can cause damage to the eyes and skin.
A
Open the front cover to go into STANDBY mode (A).
B
Unplug the electrical cord to power OFF the system (B).
Remove the Final Filter.
C
Remove the SynergyPak cartridge(C).