Maintenance
37
Installing the new UV Lamp
G
Ensure that you use the gloves supplied with the UV replacement kit.
Plug the electrical cable to the new UV Lamp (G).
H
Carefully insert the UV Lamp into the UV housing (H).
I
Attach the UV housing with the Velcro belt (I).
J
Install the SynergyPak cartridge (J).
Install the Final Filter.
Close the front cover.
Plug the electrical cord to power ON the system.
Reset the UV Timer.
(See Section How to View or Reset the UV Lamp Timer (C03))
.