Preoperative Checklist Periodic Maintenance
6 - 44 0070-10-0683 AS3000™ Service Manual
19. Test the Tidal Volume in the PCV ventilation mode:
a. Set the O
2
flow to 2 L/min and set the N
2
O and AIR flow rates to minimum flow.
b. Set the ventilator controls to:
c. Select PCV to begin ventilation.
d. Verify that the PEAK Pressure settles within ±4 of the set value within 3 breaths.
e. Re-activate CMV ventilation mode.
20. Low FiO
2
Alarm Test
a. Set the low FiO
2
Alarm limit to 50% O
2
.
b. Set the AIR flow control valve to 5 L/min.
c. Set the FiO
2
flow controller to minimum flow.
d. Verify the following Low FiO
2
alarm signals activate, within three ventilation cycles:
Low FiO
2
message appears on the screen.
An alarm tone sounds.
e. Set the Low FiO
2
alarm limit to 18%.
Verify the alarm signals stop activating.
21. High FiO
2
Alarm Test
a. Set the high FiO
2
Alarm limit to 49% O
2
.
b. Set the FiO
2
flow control valve to 5 L/min.
c. Set the AIR flow controller to minimum.
d. Verify the following High FiO
2
alarm signals activate:
High FiO
2
message appears on the screen.
An alarm tone sounds.
e. Set the high FiO
2
alarm limit to the max setting.
Verify the alarm signals stop activating.
ATTRIBUTE SETTING
Patient Type Adult
Ventilation Mode PCV
Target Pressure - P
TARGET
20
Breathng Rate - freq 8
I:E Ratio - I:E 1:2
PEEP - PEEP Off
Inspiratory Slope - Tslope 0.5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Datascope AS 3000 and is the answer not in the manual?

Mindray Datascope AS 3000 Specifications

General IconGeneral
BrandMindray
ModelDatascope AS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals