Preoperative Checklist Periodic Maintenance
6 - 46 0070-10-0683 AS3000™ Service Manual
24. Place the system in STANDBY mode.
25. Check for appropriate level of patient suction.
26. Check, connect, and calibrate other electronic monitors.
27. Turn on and set other appropriate alarms for equipment to be used.
NOTE: The following step should be performed every 3 days or
when prompted by the machine.
28. Perform the Oxygen Sensor Calibration (Refer to section 6.8.4.6 (pg. 6-
22) ‘‘Oxygen Sensor Calibration’’).
NOTE: The following step should be performed weekly or
whenever a new vaporizer is installed or when CO
2
absorbent is replaced.
29. Test for leaks in the machine and vaporizers by performing the High
Pressure Leak Test as described in step 15 of this section.