Installation Checkout Procedure Installation Guide
2 - 10 0070-10-0683 AS3000™ Service Manual
9. Test the O
2
:N
2
O ratio system (applicable to -02 units only).
a. Set all flow control knobs to minimum.
b. Set the flow of O
2
to 1 L/min.
c. Open the N
2
O flow knob and verify that the N
2
O flow will not increase higher than
3.7 L/min.
d. Return flow control knobs to minimum.
10. Perform System Test.
a. Power ON the AS3000.
b. Wait until the System Self Test is complete
c. Select Continue to prompt to the Safety Valve test
FIGURE 2-2 Startup Self Test