AS3000™ Service Manual 0070-10-0683 2 - 23
Installation Guide Installation Checkout Procedure
23. Verify that the drive gas pressure loss alarm signals activate and N
2
O
stops and AIR continues to flow.
a. Set the O
2
flow for 1 L/min.
b. Set the N
2
O flow to 1 L/min.
c. Set the AIR flow to 1 L/min.
d. Disconnect the O
2
line supply to the AS3000, and Close the O
2
cylinder.
e. Verify that the flow of N
2
O stops when O
2
is lost.
f. Verify that AIR continues to flow.
g. Verify the “O
2
Supply Failure” message appears on screen, a steady audio tone
should sound.
h. Reconnect O
2
line pressure source to the AS3000.
i. Verify that the alarms are reset.
j. Verify N
2
O flow has resumed once O
2
pressure returns.
24. Verify that the battery operation and charging icons appears.
a. Set the vent mode to Standby.
b. Disconnect AC line cord.
c. Verify that the battery icon appears on the screen and indicates the charge level of
the battery.
d. Verify that the line cord icon is flashing.
e. Verify that the flow meter backlight is illuminated during battery operation.
f. When the battery icon is in a not full state, reconnect the AC line cord.
g. Verify that the battery charge icon appears on the screen.
h. Verify that the flow meter backlight is illuminated during AC voltage operation.
25. Verify operation of the work light.
a. Turn on the work light located on the bottom side of the top panel.
b. Verify that it lights in both on positions.
26. Verify that the Auxiliary O
2
and AIR Flowmeters operates.
a. Verify an AIR flow of 15 L/min can be obtained by connecting the auxiliary AIR hose
to the pressure source and opening the flow meter.
b. Verify an O
2
flow of 15 L/min can be obtained by connecting the auxiliary O
2
hose
to the pressure source and opening the flow meter.
27. Verify that the Suction Regulator operates in LINE and REGULATE
positions.
a. Set the suction regulator's selection dial to LINE.
b. Verify maximum suction vacuum on the regulator's gauge.
c. Set the suction regulator's selection dial to REGULATE.
d. Verify the suction vacuum is adjustable on the regulator's gauge.
e. Set the suction regulator's selection dial to OFF.