Electrical Safety Inspection 
A-7 
 
ELECTRICAL SAFETY INSPECTION 
5- Earth Leakage Test 
OVERVIEW 
Run an Earth Leakage test on the device being tested before performing any other leakage
tests.
Leakagecurrentismeasuredthefollowingways:
♦EarthLeakageCurrent,leakagecurrentmeasuredthroughDUToutletEarth
♦EarthLeakageCurrentAP‐EARTH(ALLAppliedPartsconnectedtoEarth),leakagecurrent
measuredthroughDUT
outletEarth
There is no need to attach a test lead; the 601PRO automatically connects the measuring
deviceinternally.
TEST PROCEDURE 
  Perform the Test   
1)  FromtheMAINMENU,orwiththe outletunpowered,plugtheDUTinto the601PROfront
paneloutlet,andturnonthedevice.
2)  Attachthedevice'sappliedpartstothe601PROappliedpartterminalsifapplicable.
3)  Press shortcut key 4.The Earth Leakage test appears on the display, and
 the test begins
immediately:
¾  SOFTKEY1togglestheDUToutletPolarityfromNormaltoOfftoReverse.
¾  SOFTKEY2togglestheDUToutletfromEarthtoNoEarth.
¾  SOFTKEY3togglestheDUToutletfromL2toNoL2.
¾  SOFTKEY4togglestheAPtoEarth
toNoAPtoEarth.
4)  Presstheprintdatakeyatanytimetogenerateaprintout ofthelatestmeasurement.
 
  Failure
Check any broken of the AC/DC adapter and its cable. Replace a new one if any portion
defective.
Checkanybrokenoftheenclosure.Replaceanydefectivepart.
Inspectwiringforbadcrimps,poorconnections,ordamage.
Testthewalloutlet;verifyitisgroundedandisfreeofotherwiring
abnormalities.Notifythe