Electrical Safety Inspection
A-8
ELECTRICAL SAFETY INSPECTION
5- Earth Leakage Test
userorownertocorrectanydeviations.Asaworkaround,checktheotheroutletstoseeifthey
couldbeusedinstead.
Changeanotherprobetoconfirmifthefailiscausedbyconsole.
Inspectwiringforbadcrimps,poorconnections,ordamage.
If the leakage current measurement tests fail
on a new unit and if situation can not be
corrected, submit a Safety Failure Report to document the system problem. Remove unit from
operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and
disposal.
LIMITS
USA300 μA Normal Condition
1000 μA Single Fault Condition
OTHER COUNTRIES: 500 μA Normal Condition
1000 μA Single Fault Condition