6 - 48 Operator’s Manual
6 Image Acquisition
4-9 / Positioning
(probe) sensor port
Motion (abdomen)
sensor port
3. Positioning (probe) sensor: fix the positioning (probe)
sensor (2 in the figure) which is already plugged in
positioning (probe) sensor port to the cavity of the
positioning (probe) navigation support (1 in the figure)
following guide.
NOTE:
Note: if failed to plug the positioning (probe) sensor
following guide, it may lead to non-correlation of
spatial positioning.
4. Positioning (probe) sensor sterile sheath: wear the probe
sterile sheath after probe sensor and positioning (probe)
navigation support are installed.
5. Motion (abdomen) sensor: fix the motion (abdomen)
sensor which is already plugged in motion (abdomen) sensor
port to the cavity of the motion (abdomen) sensor support
following guide.
6. Motion (abdomen) sensor sterile sheath: wear the sterile
sheath after abdomen sensor and positioning (probe)
navigation support are installed.
7. Motion (abdomen) sensor support: stick the sensor
support against the top left of the patient’s abdomen with the
medical plasters (below left costal arch, right above the
navel, the area where the skin moves apparently with the
respiration).
10 Magnetic
navigator
/ Place the magnetic navigator behind the Ultrasound System.
Power on the navigator to enable the positioning function.
No. Device
Name
Port Description