2 - 2 Operator’s Manual
2 System Overview
According to the mode of operation:
CONTINUOUS OPERATION
The device equipped with the applied part of defibrillation protection:
The device is not equipped with the applied part of defibrillation protection.
According to the signal input and output parts of the device:
The device is equipped with signal input and output parts
Permanently installed or non-permanently installed:
Non-permanently installed
2.4 Product Specifications
NOTE:
The functions described in the operators manual may vary depending on the specific system
purchased.
2.4.1 Power Supply
Power input
Voltage: 220-240 V~
Frequency: 50/60 Hz
Power consumption: 1000 VA
Power output
Voltage: 220-240 V~
Frequency: 50/60 Hz
Power consumption: 300 VA (note: this is the maximum auxiliary output power of the outlet. When
the peripheral device is connected to the outlet, ensure that the maximum auxiliary output power
does not exceed this threshold. Otherwise, the system may be damaged.)
2.4.2 Environmental Conditions
WARNING
Do not use this system in conditions other than those specified.
Operating conditions
Ambient temperature: 0 °C ~ 40 °C
Relative humidity: 20% ~ 85% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
Storage and transportation conditions
Ambient temperature: -20 °C ~ 55 °C
Relative humidity: 20% ~ 95% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa